Background

Transgastric and transvaginal approaches in natural orifice transluminal endoscopic surgery (NOTES) are the most commonly used, although the transcolonic approach may have some advantages.

Objective

To develop a workable technique for transcolonic NOTES.

Design

A nonsurvival study followed by a survival study in a porcine model. Transcolonic peritoneoscopy was performed with the use of a novel intracolonic occlusion balloon. The colotomy was closed with endoclips. A necropsy and histologic evaluation were performed 2 weeks after surgery.

Setting

Academic hospital, health science research center.

Subjects

Fifteen female Yorkshire pigs (5 nonsurvival, 10 survival).

Interventions

A balloon-tipped catheter was placed proximal to the colotomy site. The balloon was inflated to occlude the colonic lumen. An endoscope was inserted through the anus. Colonic incision was created with an endoscopic needle-knife at 15 to 20 cm from the anal verge. Peritoneoscopy was performed. The colotomy was closed with endoclips.

Main Outcome Measurements

Rates of complications, survival, healing, and adhesions.

Results

Two initial experiments were complicated by bowel distension and contamination of the incision area by colonic content. In the remaining 13 pigs, the experiments were performed with the use of the intracolonic occlusion balloon. No complications were documented. Necropsies were performed 2 weeks after surgery. Gross and histologic evaluations demonstrated near complete healing. Minimal adhesions were identified in 4 of 10 pigs.

Limitation

Imperfection of the prototype balloon.

Conclusions

Excessive bowel distension and fecal contamination because of spillage from the proximal bowel may be barriers to performing transcolonic NOTES. Isolation of the operative area by splitting the bowel and sealing the colonic lumen with the balloon above the colonic incision may overcome these problems and optimize the technique.

Abbreviations: IM, intramuscular, NOTES, natural orifice transluminal endoscopic surgery

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0016-5107(08)01848-8

doi:10.1016/j.gie.2008.05.019

Gastrointestinal Endoscopy
Volume 68, Issue 4 , Pages 760-766, October 2008