Enhanced ultrasound imaging

      Abbreviations:

      ASGE (American Society for Gastrointestinal Endoscopy), CE-EUS (contrast-enhanced EUS)
      The American Society for Gastrointestinal Endoscopy (ASGE) Technology Committee provides reviews of emerging endoscopic technologies that have an impact on the practice of GI endoscopy. Evidence-based methodology is used, with MEDLINE and PubMed literature searches to identify pertinent clinical studies on the topic. Because many topics have a limited number of peer-reviewed manuscripts, abstracts from scientific meetings are used to supplement the review. The reports focus on the current status of the technologies, areas in need of further research, and barriers to incorporation into the mainstream practice of GI endoscopy.
      Reports on Emerging Technologies are drafted by one or two members of the ASGE Technology Committee, reviewed and edited by the committee as a whole, and approved by the governing board of the ASGE. These reports are scientific reviews provided solely for educational and informational purposes. Reports on Emerging Technologies are not rules and should not be construed as establishing a legal standard of care or as encouraging, advocating, requiring, or discouraging any particular treatment or payment for such treatment.

      Background

      EUS has assumed a primary role in the evaluation of intraluminal and extraluminal pathology. However, even with the addition of tissue sampling via FNA, some limitations exist in the differentiation of benign from malignant processes. Recent innovations in EUS intended to address these limitations include contrast-enhanced EUS (CE-EUS) and EUS elastography.

      Emerging technologies

      Elastography

      Elastography is a modality that evaluates tissue stiffness by its response to compression. US waves travel at different speeds through tissues of different stiffness. Compression of tissue changes its mechanical properties and its reflection of US waves (Fig. 1). An abnormal lesion may deform to a lesser degree in response to compression than surrounding normal tissue. For example, malignant tumors exhibit increased tissue stiffness compared with benign tumors or normal tissue.
      • Ophir J.
      • Caspedes I.
      • Ponnekanti H.
      • et al.
      Elastography: a quantitative method for imaging the elasticity of biological tissue.
      • Krouskop T.A.
      • Wheeler T.M.
      • Kallel F.
      • et al.
      Elastic moduli of breast and prostate tissues under compression.
      • Ophir J.
      • Alam S.K.
      • Garra B.
      • et al.
      Elastography: ultrasonic estimation and imaging of the elastic properties of tissues.
      Elastography compares the spatial arrangement of the tissue and the velocity of US waves at rest and after compression.
      • Giovannini M.
      Contrast-enhanced endoscopic ultrasound and elastosonoendoscopy.
      • Jacobson B.C.
      Pressed for an answer: has elastography finally come to EUS?.
      • Hirooka Y.
      • Itoh A.
      • Kawashima H.
      • et al.
      Diagnosis of pancreatic disorders using contrast-enhanced endoscopic ultrasonography and endoscopic elastography.
      When the tissue of a lesion is harder compared with the surrounding normal organ, the echoes will be less distorted than in the surrounding tissues.
      Figure thumbnail gr1
      Figure 1The principle of elastography. The tissue is evaluated with US. A, At rest, the harder tissue (depicted as the circular marble mass) within the softer organ. B, After compression, the harder tissue will be less distorted than the surrounding tissues. The US waves will travel faster through the less-deformed hard mass. E, echoendoscope.
      Elastography with transcutaneous US has been used in the evaluation of organs such as breast, thyroid, and liver,
      • Garra B.S.
      Imaging and estimation of tissue elasticity by ultrasound.
      and recently has become available for use with EUS.
      • Giovannini M.
      Contrast-enhanced endoscopic ultrasound and elastosonoendoscopy.
      • Jacobson B.C.
      Pressed for an answer: has elastography finally come to EUS?.
      • Hirooka Y.
      • Itoh A.
      • Kawashima H.
      • et al.
      Diagnosis of pancreatic disorders using contrast-enhanced endoscopic ultrasonography and endoscopic elastography.
      Developments in software and imaging processing (HV-900; Pentax, Montvale, NJ) allow for measurement of tissue response to both external mechanical stimuli (eg, pushing the endoscope against the esophageal wall) and/or physiologic movement (eg, vascular pulsations). Real-time elastography is performed with standard echoendoscopes (EG-3870UTK linear, EG-3670URK radial, Pentax) and displayed as an overlay of transparent colors over the regular gray-scale B-mode US image. Softer tissue is depicted as red, intermediate areas as green, and stiffer areas as blue. These color differences highlight tissue changes of malignant or fibrotic lesions compared with surrounding normal tissue.

      Potential applications

      Conventional EUS with FNA in the evaluation of solid pancreatic masses is limited by relatively low sensitivity for malignancy.
      • Varadarajulu S.
      • Tamhane A.
      • Eloubeidi M.A.
      Yield of EUS-guided FNA of pancreatic masses in the presence or the absence of chronic pancreatitis.
      • Fritscher-Ravens A.
      • Brand L.
      • Knofel W.T.
      • et al.
      Comparison of endoscopic ultrasound-guided fine needle aspiration for focal pancreatic lesions in patients with normal parenchyma and chronic pancreatitis.
      • Savides T.J.
      • Donohue M.
      • Hunt G.
      • et al.
      EUS-guided FNA diagnostic yield of malignancy in solid pancreatic masses: a benchmark for quality performance measurement.
      In an attempt to improve sensitivity, elastography has been applied in the evaluation of focal pancreatic lesions. Early studies used a color display and/or a relative numerical scoring system (score 1 to 5) to evaluate the stiffness of the tissue. They demonstrated good sensitivity but poor specificity in differentiating between malignant and benign lesions such as inflammatory or fibrotic masses.
      • Giovannini M.
      • Hookey L.C.
      • Bories E.
      • et al.
      Endoscopic ultrasound elastography: the first step towards virtual biopsy? Preliminary results in 49 patients.
      • Săftoiu A.
      • Vilman P.
      Endoscopic ultrasound elastography: a new imaging technique for the visualization of tissue elasticity distribution.
      • Janssen J.
      • Dietrich C.F.
      • Will U.
      • et al.
      Endosonographic elastography in the diagnosis of mediastinal lymph nodes.
      • Hirche T.O.
      • Ignee A.
      • Barreiros A.P.
      • et al.
      Indications and limitations of endoscopic ultrasound elastography for evaluation of focal pancreatic lesions.
      Reported limitations included difficulties with delineation of the margin of lesions and evaluation of larger (>35 mm) or deeper lesions. More recently, a multicenter series of 121 patients reported an 89% accuracy rate.
      • Giovannini M.
      • Thomas B.
      • Erwan B.
      • et al.
      Endoscopic ultrasound elastography for evaluation of lymph nodes and pancreatic masses: a multicenter study.
      A prospective, single-center study of 130 consecutive patients with pancreatic masses also revealed high accuracy (94%) for prediction of malignancy, with excellent interobserver agreement between the two endosonographers (k = 0.772).
      • Iglesias-Garcia J.
      • Larino-Noia J.
      • Abdulkader I.
      • et al.
      EUS elastography for the characterization of solid pancreatic masses.
      Disagreement occurred in 8 cases of inflammatory masses and 1 neuroendocrine tumor.
      Newer elastography software calculates the strain ratio between a lesion and adjacent softer tissue, allowing for the addition of quantitative measurement to the qualitative pattern and color recognition. In a single-center study of 86 consecutive patients with solid pancreatic masses, quantitative elastography demonstrated excellent sensitivity (100%) but lower specificity (92%) for malignancy.
      • Iglesias-Garcia J.
      • Larino-Noia J.
      • Abdulkader I.
      • et al.
      Quantitative endoscopic ultrasound elastography: an accurate method for the differentiation of solid pancreatic masses.
      Another innovation is software that analyzes elastographic images. This was developed to make image interpretation less subjective and to increase the accuracy of differentiation between benign and malignant lesions. This software had good accuracy (89%) in a feasibility study.
      • Săftoiu A.
      • Vilmann P.
      • Gorunescu F.
      • et al.
      Neural network analysis of dynamic sequences of EUS elastography used for the differential diagnosis of chronic pancreatitis and pancreatic cancer.
      Elastographic evaluation of lymph nodes also has been reported.
      • Giovannini M.
      • Hookey L.C.
      • Bories E.
      • et al.
      Endoscopic ultrasound elastography: the first step towards virtual biopsy? Preliminary results in 49 patients.
      • Săftoiu A.
      • Vilmann P.
      • Ciurea T.
      • et al.
      Dynamic analysis of endoscopic ultrasound (EUS) elastography used for the differentiation of benign and malignant lymph nodes.
      • Janssen J.
      • Dietrich C.F.
      • Will U.
      • et al.
      Endosonographic elastography in the diagnosis of mediastinal lymph nodes.
      It performed better than the standard morphologic EUS criteria for malignancy.
      • Jacobson B.C.
      Pressed for an answer: has elastography finally come to EUS?.
      Other potential applications in gastroenterology include EUS evaluation of focal liver lesions and the postradiotherapy evaluation of rectal cancer.
      • Rustemovic N.
      • Hrstic I.
      • Opacic M.
      • et al.
      EUS elastography in the diagnosis of focal liver lesions.
      • Mezzi G.
      • Arcidiacono P.G.
      • Carrara S.
      • et al.
      Elastosonography in malignant rectal disease: preliminary data.

      CE-EUS

      CE-EUS consists of injection of contrast medium during standard EUS examination. It uses the altered vascular characteristics of malignancy to enhance visualization compared with surrounding tissues.
      Current US contrast agents are composed of a microsphere containing gas bubbles within a shell (microbubbles).
      • Claudon M.
      • Cosgrove D.
      • Albrecht T.
      • et al.
      Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS)—update 2008.
      • Feinstein S.B.
      • Coll B.
      • Staub D.
      • et al.
      Contrast enhanced ultrasound imaging.
      After intravenous administration, the US contrast medium remains in the vascular space and enhances the visualization of blood vessels, from large vessels and heart chambers to the microcirculation within the pancreatic parenchyma. For the pancreas, an initial macrovascular phase followed by a microvascular phase has been described.
      • Fusaroli P.
      • Spada A.
      • Mancino M.G.
      • et al.
      Contrast harmonic echo-endoscopic ultrasound improves accuracy in diagnosis of solid pancreatic masses.
      • Napoleon B.
      • Alvarez-Sanchez M.V.
      • Gincoul R.
      • et al.
      Contrast-enhanced harmonic endoscopic ultrasound in solid lesions of the pancreas: results of a pilot study.
      The contrast reaches maximum intensity in 20 to 30 seconds after injection but may last up to 5 to 10 minutes in the microcirculation.
      • D'Onofrio M.
      • Zamboni G.
      • Tognolini M.
      • et al.
      Ultrasonography of the pancreas: 4-contrast-enhanced imaging.
      Microbubble agents available in the US contain perfluoropropane gas microspheres and are marketed as Optison (GE Healthcare, Princeton, NJ), approved in 1997, and Definity (Lantheus Medical Imaging, Billerica, Mass), approved in 2001. In October 2007, the U.S. Food and Drug Administration issued a black box warning and new contraindications for contrast agents containing perfluoropropane because of reports of cardiac ischemia. However, some of these cardiac restrictions were recently eased.
      Information for healthcare professionals: Micro-bubble contrast agents (marketed as Definity (perflutren lipid microsphere) injectable suspension and Optison (perflutren protein-type A microspheres for injection).
      Overall, adverse event rates are low (0.6%) and include mainly back pain and headache.
      • Agelli C.
      • Giannopoulous G.
      • Roussakis G.
      • et al.
      Safety of myocardial flash-contrast echocardiography in combination with dobutamine stress testing for the detection of ischaemia in 5250 studies.
      • Khawaja O.A.
      • Shaikh K.A.
      • Al-Mallah M.H.
      Meta-analysis of adverse cardiovascular events associated with echocardiographic contrast agents.
      Hypersensitivity to perfluoropropane is rare (0.014%) but can be severe, including anaphylactoid or anaphylactic reactions.
      • Wilson S.R.
      • Burns P.N.
      Microbubble-enhanced US in body imaging: what role?.
      Color Doppler or power Doppler was the initial EUS mode used for CE-EUS. This resulted in improved imaging compared with standard B-mode, but this had a low sensitivity for microvasculature. The resulting image was blemished with motion artifacts and blooming effect in blood vessels.
      • Dietrich C.F.
      • Ignee A.
      • Braden B.
      • et al.
      Improved differentiation of pancreatic tumors using contrast-enhanced endoscopic ultrasound.
      • Hocke M.
      • Schulze E.
      • Gottschalk P.
      • et al.
      Contrast-enhanced endoscopic ultrasound in discrimination between focal pancreatitis and pancreatic cancer.
      Newer echoendoscopes with wider bandwidth transducers and processors with harmonic imaging software (GF-UE160-AL5, ExPHD detection mode in Aloka Prosound SSD Alpha-10, Olympus) may improve resolution from areas of low blood flow when used with contrast medium.
      • Dietrich C.F.
      Contrast-enhanced low mechanical index endoscopic ultrasound (CELMI-EUS).

      Potential applications

      Several clinical studies have evaluated CE-EUS with color Doppler or power Doppler. Hypovascularity during contrast enhancement (hypoenhancing lesion) was typically associated with malignancy and other lesions, such as neuroendocrine tumors, were hypervascular (hyperenhancing lesion) or isovascular relative to the normal parenchyma.
      • Wilson S.R.
      • Burns P.N.
      Microbubble-enhanced US in body imaging: what role?.
      • Rickes S.
      • Monkemuller K.
      • Malfertheiner P.
      Contrast-enhanced ultra-sound in the diagnosis of pancreatic tumors.
      Small pancreatic endocrine tumors were more clearly seen with contrast-enhanced color Doppler EUS in a retrospective series of 41 patients with 62 lesions. When compared with pathologic confirmation, EUS with contrast had an excellent sensitivity of 95%.
      • Ishikawa T.
      • Itoh A.
      • Kawashima H.
      • et al.
      Usefulness of EUS combined with contrast-enhancement in the differential diagnosis of malignant versus benign and preoperative localization of pancreatic endocrine tumors.
      CE-EUS using harmonic imaging has been used in the evaluation of intra-abdominal lesions of undetermined etiology with high accuracy (97%) for malignancy.
      • Xia Y.
      • Kitano M.
      • Kudo M.
      • et al.
      Characterization of intra-abdominal lesions of undetermined origin by contrast-enhanced harmonic EUS (with videos).
      Another study of solid pancreatic masses showed excellent sensitivity (96%) and good accuracy (82%) for malignancy for CE-EUS using harmonic imaging mode.
      • Fusaroli P.
      • Spada A.
      • Mancino M.G.
      • et al.
      Contrast harmonic echo-endoscopic ultrasound improves accuracy in diagnosis of solid pancreatic masses.
      A noteworthy finding in this study was the detection of small lesions in seven patients with biliary stents or chronic pancreatitis that were not initially seen with standard EUS images, leading to an 8% increase in overall diagnostic yield. Two other small studies of pancreaticobiliary malignancy also showed good sensitivity and accuracy.
      • Napoleon B.
      • Alvarez-Sanchez M.V.
      • Gincoul R.
      • et al.
      Contrast-enhanced harmonic endoscopic ultrasound in solid lesions of the pancreas: results of a pilot study.
      • Imazu H.
      • Uchiyama Y.
      • Matsunagna K.
      • et al.
      Contrast-enhanced harmonic EUS with novel ultrasonographic contrast (Sonazoid) in the preoperative T-staging for pancreaticobiliary malignancies.
      A single-center report suggests that the combination of both elastography and CE-EUS may identify patients with focal pancreatic masses that need a more aggressive approach with repeat EUS-FNA or even referral for surgery. In 54 patients with focal pancreatic masses, the combination had a very good positive predictive value (96%) and negative predictive value of 71%.
      • Săftoiu A.
      • Iordache S.A.
      • Gheonea D.I.
      • et al.
      Combined contrast-enhanced power Doppler and real-time sonoelastography performed during EUS, used in the differential diagnosis of focal pancreatic masses (with videos).

      Areas for future research

      Multicenter prospective studies comparing CE-EUS and elastography to other diagnostic techniques are needed. Specifically, the role of these technologies in differentiating between benign and malignant pathology needs to be further clarified. Validation and cost-effectiveness studies also are warranted. Additionally, prospective data may help establish cost-effectiveness. Newer microbubbles that specifically target certain types of malignancy on a molecular basis are being developed.
      • Kiessling F.
      Science to practice: the dawn of molecular US imaging for clinical cancer imaging.
      The efficacy and safety of these target-specific agents await further study.

      Summary

      Elastography and CE-EUS are new technologies that may improve the real-time evaluation of extraluminal lesions. Further studies are needed to define their role.

      Disclosure

      No financial relationships relevant to this publication were disclosed.

      References

        • Ophir J.
        • Caspedes I.
        • Ponnekanti H.
        • et al.
        Elastography: a quantitative method for imaging the elasticity of biological tissue.
        Ultrason Imaging. 1991; 13: 111-134
        • Krouskop T.A.
        • Wheeler T.M.
        • Kallel F.
        • et al.
        Elastic moduli of breast and prostate tissues under compression.
        Ultrason Imaging. 1998; 20: 260-274
        • Ophir J.
        • Alam S.K.
        • Garra B.
        • et al.
        Elastography: ultrasonic estimation and imaging of the elastic properties of tissues.
        Proc Inst Mech Eng H. 1999; 213: 203-233
        • Giovannini M.
        Contrast-enhanced endoscopic ultrasound and elastosonoendoscopy.
        Best Pract Res Clin Gastroenterol. 2009; 23: 767-779
        • Jacobson B.C.
        Pressed for an answer: has elastography finally come to EUS?.
        Gastrointest Endosc. 2007; 66: 301-303
        • Hirooka Y.
        • Itoh A.
        • Kawashima H.
        • et al.
        Diagnosis of pancreatic disorders using contrast-enhanced endoscopic ultrasonography and endoscopic elastography.
        Clin Gastroenterol Hepatol. 2009; 7: S63-S67
        • Garra B.S.
        Imaging and estimation of tissue elasticity by ultrasound.
        [Review] Ultrasound Q. 2007; 23: 255-268
        • Varadarajulu S.
        • Tamhane A.
        • Eloubeidi M.A.
        Yield of EUS-guided FNA of pancreatic masses in the presence or the absence of chronic pancreatitis.
        Gastrointest Endosc. 2005; 62: 728-736
        • Fritscher-Ravens A.
        • Brand L.
        • Knofel W.T.
        • et al.
        Comparison of endoscopic ultrasound-guided fine needle aspiration for focal pancreatic lesions in patients with normal parenchyma and chronic pancreatitis.
        Am J Gastroenterol. 2002; 97: 2768-2775
        • Savides T.J.
        • Donohue M.
        • Hunt G.
        • et al.
        EUS-guided FNA diagnostic yield of malignancy in solid pancreatic masses: a benchmark for quality performance measurement.
        Gastrointest Endosc. 2007; 66: 277-282
        • Giovannini M.
        • Hookey L.C.
        • Bories E.
        • et al.
        Endoscopic ultrasound elastography: the first step towards virtual biopsy?.
        Endoscopy. 2006; 38: 344-348
        • Săftoiu A.
        • Vilman P.
        Endoscopic ultrasound elastography: a new imaging technique for the visualization of tissue elasticity distribution.
        J Gastrointest Liver Dis. 2006; 15: 161-165
        • Janssen J.
        • Dietrich C.F.
        • Will U.
        • et al.
        Endosonographic elastography in the diagnosis of mediastinal lymph nodes.
        Endoscopy. 2007; 39: 952-957
        • Hirche T.O.
        • Ignee A.
        • Barreiros A.P.
        • et al.
        Indications and limitations of endoscopic ultrasound elastography for evaluation of focal pancreatic lesions.
        Endoscopy. 2008; 40: 910-917
        • Giovannini M.
        • Thomas B.
        • Erwan B.
        • et al.
        Endoscopic ultrasound elastography for evaluation of lymph nodes and pancreatic masses: a multicenter study.
        World J Gastroenterol. 2009; 15: 1587-1593
        • Iglesias-Garcia J.
        • Larino-Noia J.
        • Abdulkader I.
        • et al.
        EUS elastography for the characterization of solid pancreatic masses.
        Gastrointest Endosc. 2009; 70: 1101-1108
        • Iglesias-Garcia J.
        • Larino-Noia J.
        • Abdulkader I.
        • et al.
        Quantitative endoscopic ultrasound elastography: an accurate method for the differentiation of solid pancreatic masses.
        Gastroenterology. 2010; 139: 1172-1180
        • Săftoiu A.
        • Vilmann P.
        • Gorunescu F.
        • et al.
        Neural network analysis of dynamic sequences of EUS elastography used for the differential diagnosis of chronic pancreatitis and pancreatic cancer.
        Gastrointest Endosc. 2008; 68: 1086-1094
        • Săftoiu A.
        • Vilmann P.
        • Ciurea T.
        • et al.
        Dynamic analysis of endoscopic ultrasound (EUS) elastography used for the differentiation of benign and malignant lymph nodes.
        Gastrointest Endosc. 2007; 66: 291-300
        • Janssen J.
        • Dietrich C.F.
        • Will U.
        • et al.
        Endosonographic elastography in the diagnosis of mediastinal lymph nodes.
        Endoscopy. 2007; 39: 952-957
        • Rustemovic N.
        • Hrstic I.
        • Opacic M.
        • et al.
        EUS elastography in the diagnosis of focal liver lesions.
        Gastrointest Endosc. 2007; 66: 823-824
        • Mezzi G.
        • Arcidiacono P.G.
        • Carrara S.
        • et al.
        Elastosonography in malignant rectal disease: preliminary data.
        Endoscopy. 2007; 39: 375
        • Claudon M.
        • Cosgrove D.
        • Albrecht T.
        • et al.
        Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS)—update 2008.
        Ultraschall Med. 2008; 29: 28-44
        • Feinstein S.B.
        • Coll B.
        • Staub D.
        • et al.
        Contrast enhanced ultrasound imaging.
        J Nucl Cardiol. 2010; 17: 106-115
        • Fusaroli P.
        • Spada A.
        • Mancino M.G.
        • et al.
        Contrast harmonic echo-endoscopic ultrasound improves accuracy in diagnosis of solid pancreatic masses.
        Clin Gastroenterol Hepatol. 2010; 8: 629-634
        • Napoleon B.
        • Alvarez-Sanchez M.V.
        • Gincoul R.
        • et al.
        Contrast-enhanced harmonic endoscopic ultrasound in solid lesions of the pancreas: results of a pilot study.
        Endoscopy. 2010; 42: 564-570
        • D'Onofrio M.
        • Zamboni G.
        • Tognolini M.
        • et al.
        Ultrasonography of the pancreas: 4-contrast-enhanced imaging.
        Abdom Imaging. 2007; 32: 171-181
      1. Information for healthcare professionals: Micro-bubble contrast agents (marketed as Definity (perflutren lipid microsphere) injectable suspension and Optison (perflutren protein-type A microspheres for injection).
        (Accessed March 2, 2011)
        • Agelli C.
        • Giannopoulous G.
        • Roussakis G.
        • et al.
        Safety of myocardial flash-contrast echocardiography in combination with dobutamine stress testing for the detection of ischaemia in 5250 studies.
        Heart. 2008; 94: 1571-1577
        • Khawaja O.A.
        • Shaikh K.A.
        • Al-Mallah M.H.
        Meta-analysis of adverse cardiovascular events associated with echocardiographic contrast agents.
        Am J Cardiol. 2010; 106: 742-747
        • Wilson S.R.
        • Burns P.N.
        Microbubble-enhanced US in body imaging: what role?.
        Radiology. 2010; 257: 24-39
        • Dietrich C.F.
        • Ignee A.
        • Braden B.
        • et al.
        Improved differentiation of pancreatic tumors using contrast-enhanced endoscopic ultrasound.
        Clin Gastroenterol Hepatol. 2008; 6: 590-597
        • Hocke M.
        • Schulze E.
        • Gottschalk P.
        • et al.
        Contrast-enhanced endoscopic ultrasound in discrimination between focal pancreatitis and pancreatic cancer.
        World J Gastroenterol. 2006; 12: 246-250
        • Dietrich C.F.
        Contrast-enhanced low mechanical index endoscopic ultrasound (CELMI-EUS).
        Endoscopy. 2009; 41: E43-E44
        • Rickes S.
        • Monkemuller K.
        • Malfertheiner P.
        Contrast-enhanced ultra-sound in the diagnosis of pancreatic tumors.
        J Pancreas. 2006; 7: 584-592
        • Ishikawa T.
        • Itoh A.
        • Kawashima H.
        • et al.
        Usefulness of EUS combined with contrast-enhancement in the differential diagnosis of malignant versus benign and preoperative localization of pancreatic endocrine tumors.
        Gastrointest Endosc. 2010; 71: 951-959
        • Xia Y.
        • Kitano M.
        • Kudo M.
        • et al.
        Characterization of intra-abdominal lesions of undetermined origin by contrast-enhanced harmonic EUS (with videos).
        Gastrointest Endosc. 2010; 72: 637-642
        • Imazu H.
        • Uchiyama Y.
        • Matsunagna K.
        • et al.
        Contrast-enhanced harmonic EUS with novel ultrasonographic contrast (Sonazoid) in the preoperative T-staging for pancreaticobiliary malignancies.
        Scand J Gastroenterol. 2010; 45: 732-738
        • Săftoiu A.
        • Iordache S.A.
        • Gheonea D.I.
        • et al.
        Combined contrast-enhanced power Doppler and real-time sonoelastography performed during EUS, used in the differential diagnosis of focal pancreatic masses (with videos).
        Gastrointest Endosc. 2010; 72: 739-747
        • Kiessling F.
        Science to practice: the dawn of molecular US imaging for clinical cancer imaging.
        Radiology. 2010; 256: 331-333

      Linked Article

      • Erratum
        Gastrointestinal EndoscopyVol. 74Issue 1
        • Preview
          The author list for “Enhanced ultrasound imaging”(Gstrointest Endosc 2011;73:857-60) should read in this order: Marcos C. Pedrosa, MD, MPH, Bradley A. Barth, MD, NASPGHAN Representative, David J. Desilets, MD, Vivek Kaul, MD, Sripathi R. Kethu, MD, Patrick R. Pfau, MD, Jeffrey L. Tokar, MD, Shyam Varadarajulu, MD, Amy Wang, MD, Louis-Michel Wong Kee Song, MD, Sarah A. Rodriguez, MD, Committee Chair.
        • Full-Text
        • PDF