Advertisement

A quarantine process for the resolution of duodenoscope-associated transmission of multidrug-resistant Escherichia coli

      Background

      Because of their complex design, duodenoscopes have been long recognized to be difficult to fully disinfect and may play a role in transmission of bacteria between patients. Recent reports of duodenoscope-associated carbapenem-resistant enterobacteriaceae transmission have confirmed these suspicions. An outbreak of a multidrug resistant strain of Escherichia coli was recently reported at our institution. Herein we report the results of our investigation and the process improvements that we deployed in an effort to contain the outbreak.

      Methods

      A full investigation into the environment, endoscopists, infection control practices, high-level disinfection process as well as endoscopes was undertaken in conjunction with the local county health authority and the Centers for Disease Control and Prevention. Duodenoscopes were cultured and quarantined for 48 hours until negative cultures were obtained. Ergonomic changes were made to the endoscope reprocessing area, duodenoscopes were returned for routine maintenance, and surveillance cultures were obtained from all patients undergoing ERCP.

      Results

      Between November 2012 and August 2013, 32 patients were found to harbor 1 of 2 clonal strains of multidrug-resistant E coli, all of whom had undergone ERCP or duodenoscopy. A total of 1149 ERCPs were performed during this time period. Seven patients died within 31 days of the organism being identified in culture, 16 patients died overall by March 2015. The exact contribution of E coli to death is unclear because most patients had underlying late-stage malignancy or other severe medical comorbidities. No breach in high-level disinfection protocol or infection control practices was identified. The clonal strain of E coli was identified in culture on 4 of 8 duodenoscopes, 3 of which required critical repairs despite lack of obvious malfunction. The defect rate in high-level disinfection of duodenoscopes was 2% over a 1-year period. The implemented quality improvements, subsequent to which 1625 ERCPs have been performed, were successful in halting the outbreak.

      Conclusions

      The existing manufacturer-recommended high-level disinfection protocols for duodenoscopes are inadequate. Although the ultimate solution may be a design change to the instrument, the timeline for such a change appears long and potentially difficult to exact. In the interim, a reliable method to ensure that bacterial pathogens are not present on the duodenoscope after high-level disinfection is needed.

      Abbreviations:

      HAC (hyper AmpC), HLD (high-level disinfection), MDRO (multidrug resistant organisms)
      To read this article in full you will need to make a payment

      Subscribe:

      Subscribe to Gastrointestinal Endoscopy
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Bond W.W.
        Overview of infection control problems: principles in gastrointestinal endoscopy.
        Gastrointest Endosc Clin N Am. 2000; 10: 199-213
        • Spach D.
        • Silverstein F.
        • Stamm W.
        Transmission of infection by gastrointestinal endoscopy and bronchoscopy.
        Ann Intern Med. 1993; 118: 117-128
        • Aumeran C.
        • Poincloux L.
        • Souweine B.
        • et al.
        Multidrug-resistant Klebsiella pneumonia outbreak after endoscopic retrograde cholangiopancreatography.
        Endoscopy. 2010; 42: 895-899
        • Gastmeier P.
        • Vonberg R.P.
        Klebsiella spp. in endoscopy-associated infections: we may only be seeing the tip of the iceberg.
        Infection. 2014; 42: 15-21
        • Alrabaa S.F.
        • Nguyen P.
        • Sanderson R.
        • et al.
        Early identification and control of carbapenemase-producing Klebsiella pneumoniae, originating from contaminated endoscopic equipment.
        Am J Infect Control. 2013; 41: 562-564
        • Petersen B.T.
        • Chennat J.
        • Cohen J.
        • et al.
        Multisociety guideline on reprocessing flexible gastrointestinal endoscopes: 2011.
        Gastrointest Endosc. 2011; 73: 1075-1084
        • Epstein L.
        • Hunter J.C.
        • Arwady M.A.
        • et al.
        New Delhi metallo-b-lactamase producing carbapenem-resistant Escherichia coli associated with exposure to duodenoscopes.
        JAMA. 2014; 312: 1447-1455
        • Smith Z.
        • Oh Y.
        • Saeian K.
        • et al.
        Transmission of carbapenem-resistant Enterobacteriaceae during ERCP: time to revisit the current reprocessing guidelines.
        Gastrointest Endosc. 2015; 81: 1041-1045
      1. Wendorf KA, MD, Kay M, Baliga C. Endoscope-associated multidrug-resistant Escherichia coli outbreak — King County, Washington, 2012–2014. Abstract presented at ID Week, Philadelphia (Pa); October 2014.

        • Wendorf K.
        • Kay M.
        • Baliga C.
        • et al.
        Endoscopic retrograde cholangiopancreatography-associated AmpC Escherichia coli outbreak.
        Infect Control Hosp Epidemiol. 2015; Mar 30: 1-9
        • Furman C.
        • Kaplan R.
        Applying the Toyota Production System: using a patient safety alert system to reduce error.
        Jt Comm J Qual Patient Saf. 2007; 33: 376-386
        • Low D.E.
        • Micflikier A.B.
        • Kennedy J.K.
        • et al.
        Infectious complications of endoscopic retrograde cholangiopancreatography: a prospective assessment.
        Arch Intern Med. 1980; 140: 1076-1077
        • Nelson D.
        • Barkun A.
        • Block K.
        • et al.
        Transmission of infection by gastrointestinal endoscopy.
        Gastrointest Endosc. 2001; 54: 824-828
        • Doherty D.E.
        • Falko J.M.
        • Lefkovitz N.
        • et al.
        Pseudomonas aeruginosa sepsis following endoscopic retrograde cholangiopancreatography (ERCP).
        Dig Dis Sci. 1982; 27: 169-170
        • Earnshaw J.J.
        • Clark A.W.
        • Thom B.T.
        Outbreak of Pseudomonas aeruginosa following endoscopic retrograde cholangiopancreatography.
        J Hosp Infect. 1985; 6: 95-97
        • Allen J.I.
        • Allen M.O.
        • Olson M.M.
        • et al.
        Pseudomonas infection of the biliary system resulting from use of a contaminated endoscope.
        Gastroenterology. 1987; 92: 759-763
      2. Design of endoscopic retrograde cholangiopancreatography (ERCP) duodenoscopes may impede effective cleaning: FDA safety communication. Issued February 19th, 2015. Available at: http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm434871.htm. Accessed February 19, 2015.

        • Haney P.E.
        • Raymond B.A.
        • Lewis L.C.
        Ethylene oxide. An occupational health hazard for hospital workers.
        AORN J. 1990; 51: 480-486
      3. Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.
        Clin Microbiol Rev. 2013; 26: 231-254
        • O’Conner B.H.
        • Bennet J.R.
        • Alexander J.F.
        • et al.
        Salmonellosis infection transmitted by fiberoptic endoscopes.
        Lancet. 1982; 2: 864-866
        • Dwyer D.M.
        • Klein E.G.
        • Istre G.R.
        • et al.
        Salmonella newport infections transmitted by fiberoptic colonoscopy.
        Gastrointest Endosc. 1987; 33: 84-87
      4. Interim duodenoscope surveillance protocol. United States Centers for Disease Control and Prevention. Released March 11, 2015. Available at: http://www.cdc.gov/hai/organisms/cre/cre-duodenoscope-surveillance-protocol.html. Accessed March 11, 2015.