Novel single-use duodenoscope compared with 3 models of reusable duodenoscopes for ERCP: a randomized bench-model comparison

Published:October 31, 2019DOI:https://doi.org/10.1016/j.gie.2019.08.032

      Background and Aims

      Multidrug-resistant infectious outbreaks associated with duodenoscope reuse have been documented internationally. A single-use endoscope could eliminate exogenous patient-to-patient infection associated with ERCP.

      Methods

      We conducted a comparative bench simulation study of a new single-use and 3 models of reusable duodenoscopes on a synthetic anatomic bench model. Four ERCP tasks were performed: guidewire locking (single-use and 1 reusable duodenoscope only), plastic stent placement and removal, metal stent placement and removal, and basket sweeping. The study schedule included block randomization by 4 duodenoscopes, 4 tasks, and 2 anatomic model ERCP stations. Ability to complete tasks, task completion times, and subjective ratings of overall performance, navigation/pushability, tip control, and image quality on a scale of 1 (worst) to 10 (best) were compared among duodenoscopes.

      Results

      All 4 ERCP tasks (total 14 subtasks) were completed by 6 expert endoscopists using all 4 duodenoscopes, with similar task completion times (median, 1.5-8.0 minutes per task) and overall performance ratings by task (median, 8.0-10.0). Navigation/pushability ratings were lower for the single-use duodenoscope than for the 3 reusable duodenoscopes (median, 8.0, 10.0, 9.0, and 9.0, respectively; P < .01). Tip control ratings were similar among all the duodenoscopes (median, 9.0-10.0; P = .77). Image quality ratings were lower for 1 reusable duodenoscope compared with the single-use and other 2 reusable duodenoscopes (median, 8.0, 9.0, 9.0, and 9.0, respectively; P < .01).

      Conclusions

      A new single-use duodenoscope was used to simulate 4 ERCP tasks in an anatomic model, with performance ratings and completion times comparable with 3 models of reusable duodenoscopes.
      To read this article in full you will need to make a payment

      Subscribe:

      Subscribe to Gastrointestinal Endoscopy
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • International Committee of Medical Journal Editors
        Defining the role of authors and contributors: 2. Who is an author?.
        (Available at:)
        • Battisti W.P.
        • Wager E.
        • Baltzer L.
        • et al.
        Good publication practice for communicating company-sponsored medical research: GPP3.
        Ann Intern Med. 2015; 163: 461-464
        • Lee D.H.
        • Kim D.B.
        • Kim H.Y.
        • et al.
        Increasing potential risks of contamination from repetitive use of endoscope.
        Am J Infect Control. 2015; 43: e13-e17
        • Rahman M.R.
        • Perisetti A.
        • Coman R.
        • et al.
        Duodenoscope-associated infections: update on an emerging problem.
        Dig Dis Sci. 2019; 64: 1409-1418
        • Petersen B.T.
        • Koch J.
        • Ginsberg G.G.
        Infection using ERCP endoscopes.
        Gastroenterology. 2016; 151: 46-50
        • Rutala W.A.
        • Weber D.J.
        Outbreaks of carbapenem-resistant Enterobacteriaceae infections associated with duodenoscopes: What can we do to prevent infections?.
        Am J Infect Control. 2016; 44: e47-e51
        • Beilenhoff U.
        • Biering H.
        • Blum R.
        • et al.
        Prevention of multidrug-resistant infections from contaminated duodenoscopes: position statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA).
        Endoscopy. 2017; 49: 1098-1106
        • Alfa M.J.
        • Singh H.
        • Duerksen D.R.
        • et al.
        Improper positioning of the elevator lever of duodenoscopes may lead to sequestered bacteria that survive disinfection by automated endoscope reprocessors.
        Am J Infect Control. 2018; 46: 73-75
        • Kovaleva J.
        • Peters F.T.
        • van der Mei H.C.
        • et al.
        Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy.
        Clin Microbiol Rev. 2013; 26: 231-254
        • Thaker A.M.
        • Kim S.
        • Sedarat A.
        • et al.
        Inspection of endoscope instrument channels after reprocessing using a prototype borescope.
        Gastrointest Endosc. 2018; 88: 612-619
        • Cassini A.
        • Hogberg L.D.
        • Plachouras D.
        • et al.
        Attributable deaths and disability-adjusted life-years caused by infections with antibiotic-resistant bacteria in the EU and the European Economic Area in 2015: a population-level modelling analysis.
        Lancet Infect Dis. 2019; 19: 56-66
        • Rauwers A.W.
        • Voor In't Holt A.F.
        • Buijs J.G.
        • et al.
        High prevalence rate of digestive tract bacteria in duodenoscopes: a nationwide study.
        Gut. 2018; 67: 1637-1645
        • McCafferty C.E.
        • Aghajani M.J.
        • Abi-Hanna D.
        • et al.
        An update on gastrointestinal endoscopy-associated infections and their contributing factors.
        Ann Clin Microbiol Antimicrob. 2018; 17: 36
        • Beilenhoff U.
        • Biering H.
        • Blum R.
        • et al.
        Reprocessing of flexible endoscopes and endoscopic accessories used in gastrointestinal endoscopy: position statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA)—update 2018.
        Endoscopy. 2018; 50: 1205-1234
        • Carbonne A.
        • Thiolet J.M.
        • Fournier S.
        • et al.
        Control of a multi-hospital outbreak of KPC-producing Klebsiella pneumoniae type 2 in France, September to October 2009.
        Eur Surveill. 2010; 15: 19734
        • Giakoupi P.
        • Maltezou H.
        • Polemis M.
        • et al.
        KPC-2-producing Klebsiella pneumoniae infections in Greek hospitals are mainly due to a hyperepidemic clone.
        Eur Surveill. 2009; 14
        • Gauthier L.
        • Dortet L.
        • Cotellon G.
        • et al.
        Diversity of carbapenemase-producing Escherichia coli isolates in France in 2012-2013.
        Antimicrob Agents Chemother. 2018; (e00266-18): 62
        • U.S. Food and Drug Administration
        Medical Devices; Medical Device Safety; Safety Communications. The FDA provides interim results of duodenoscope reprocessing studies conducted in real-world settings: FDA safety communication (December 10, 2018).
        (Available at:)
        • U.S. Food and Drug Administration
        Medical Devices; Medical Device Safety; Safety Communications. The FDA continues to remind facilities of the importance of following duodenoscope reprocessing instructions: FDA safety communication (April 12, 2019).
        (Available at:)
        • Chan J.K.
        • Ng I.
        • Ang J.P.
        • et al.
        Randomised controlled trial comparing the Ambu(R) aScope2 with a conventional fibreoptic bronchoscope in orotracheal intubation of anaesthetised adult patients.
        Anaesth Intensive Care. 2015; 43: 479-484
        • Terjesen C.L.
        • Kovaleva J.
        • Ehlers L.
        Early assessment of the likely cost effectiveness of single-use flexible video bronchoscopes.
        Pharmacoecon Open. 2017; 1: 133-141
        • Mager R.
        • Kurosch M.
        • Hofner T.
        • et al.
        Clinical outcomes and costs of reusable and single-use flexible ureterorenoscopes: a prospective cohort study.
        Urolithiasis. 2018; 46: 587-593
        • Hennessey D.B.
        • Fojecki G.L.
        • Papa N.P.
        • et al.
        Single-use disposable digital flexible ureteroscopes: an ex vivo assessment and cost analysis.
        BJU Int. 2018; 121: 55-61
        • Kozarek R.A.
        The future of ERCP.
        Endosc Int Open. 2017; 5: E272-E274
      1. NIH state-of-the-science statement on endoscopic retrograde cholangiopancreatography (ERCP) for diagnosis and therapy.
        NIH Consens State Sci Statements. 2002; 19: 1-26
        • Rauwers A.W.
        • Troelstra A.
        • Fluit A.C.
        • et al.
        Independent root cause analysis of contributing factors, including dismantling of 2 duodenoscopes, to an outbreak of multidrug-resistant Klebsiella pneumoniae.
        Gastrointest Endosc. 2019; 90: 793-804
        • Thaker A.M.
        • Muthusamy V.R.
        • Sedarat A.
        • et al.
        Duodenoscope reprocessing practice patterns in U.S. endoscopy centers: a survey study.
        Gastrointest Endosc. 2018; 88: 316-322
        • Petersen B.T.
        • Cohen J.
        • et al.
        • Reprocessing Guideline Task Force
        Multisociety guideline on reprocessing flexible GI endoscopes: 2016 update.
        Gastrointest Endosc. 2017; 85: 282-294
        • Bang J.Y.
        • Sutton B.
        • Hawes R.
        • et al.
        Concept of disposable duodenoscope: at what cost?.
        Gut. Epub. 2019 Feb 12;